one deep breath per 70 ~ 120 controlled respirations, the inspiration time is 1.5 times of the setting point
Maximum safety pressure:
≤ 12.5 kPa
Pressure limit range:
0 ~ 6.0 kPa
Airway pressure alarm: Audible and visual and with yellow and red color indicating
Lower: 0.2kPa~5.0kPa; Upper: 0.3~6.0 kPa
±0.2 kPa
Tidal volume alarm: Audible and visual and with yellow and red color indicating
upper alarm: 50 to 2000ml, lower alarm: 0~1800ml
Oxygen concentration alarm: Audible and visual and with yellow and red color indicating
upper alarm: 21% ~100%; lower alarm:10%~80%
Power Supply Alarm
Ac/dc power supply are after failing to send out alarm immediately Alarm Time: keep >120s
The airway pressure continues to be higher than 15 hPa±1 hPa for 15s±1s, then the machine will raise an audible alarm, the pressure will be displayed in red and the continuous high pressure red alarm wording is displayed on the screen of the anesthetic respirator.